|

The Science of
Microspectroscopy
Microspectrophotometer
Design
Uses of
Microspectrophotometers
Units of UV-Visible-NIR
Microspectroscopy
Algorithms of UV-Visible-NIR
Microspectroscopy
QDI
2010™ Microspectrophotometer

|
 |

The UV-visible-NIR
microspectrophotometer is
designed to measure spectra
of microscopic areas or
microscopic areas on larger samples.
They
can be configured to
measure the transmittance,
absorbance, reflectance,
polarization and
fluorescence of sample areas
as smaller than a micron.
Microspectrophotometer
Operation
The microscope
spectrometer and
microspectrometer work like this: a lamp
on the microscope emits white light which
is focused onto the sample.
The sample absorbs some
wavelengths of light better
than others...it all depends upon
the sample chromophore's chemical structure and
environment. The
light not absorbed is
collected by the microspectrophotometer
objective and focused onto
the entrance aperture of the
spectrophotometer. As
the aperture is mirrored, the
majority of light is
reflected
into a digital imaging
system.
This allows you to see the
spectrophotometer aperture
overlaid on the sample and
makes it very easy to
align the system
and take spectra. The figure
of fiber to the left shows
the entrance aperture as a
black square. Simply
move the stage so that the
square is over the sample to
make measurements.
The light that is not
reflected into the digital
imaging system will pass
through the
aperture
into the spectrophotometer.
The light is separated into
component
wavelengths by an optical
grating and each
component's
intensity is measured by a
pixel on a Charge Coupled
Device or CCD detector. The
computer stores this
information and the result
is an optical spectrum.
This spectrum is plotted as
an XY chart that shows the
optical energy at each
wavelength (see Figure of
spectra to left).
Different types of
microspectroscopy are
accomplished by different
lighting techniques.
These techniques are
determined by the samples
themselves. For
example, incident or
reflectance illumination is
used for opaque samples
whereas transmitted light is
used for transparent
samples. A
microspectrophotometer can
be configured to measure
transmission, absorbance,
reflectance and emission
spectra.
In operation, a measurement
is straightforward.
One first takes a dark scan
to measure the dark counts
of the system The
spectrum from a reference
material is then collected.
The reference spectrum
contains the spectral
characteristics of the
reference material itself,
the light sources, the
optics and the CCD.
The spectrum from the sample
is then acquired and an
algorithm is used
to calculate the appropriate
spectra for that lighting
condition i.e. reflectance
spectra when measuring
incident illumination.
The algorithm is
automatically applied by the
computer and the result is
displayed as a spectrum.
|